Instruction Manual CON 700 Conductivity/ºC/ºF Bench Meter TechnologyMadeEasy... Part of Thermo Fisher Scientific
The factory default is Single-Point Calibration. See Section 6.11 to change this setting. 4.3 General Calibration Tips For best results, periodic cal
4.5 Manual Conductivity & TDS Calibration Procedure 1) Press as needed to select conductivity (μS or mS) or TDS (ppm or ppt) calibration. 2) D
2) Dip the probe into the sample beyond the upper steel band (utilize the fill line on the outside of the probe guard for reference). 3) Allow time
2) To recall data from memory, press . The location of the most recent stored data is displayed first. Press or to select the location of the des
Press to confirm. 6.5 P3. (C hrenheit) 2 ºC ºF elcius Or FaPress . Press or to select ºC or ºF. Press to3 AtC (Automatic Tem confirm. 6.6
Press . Press or to select the desired value (0.00 to 10.0). to confirm. Press 6.9 P3.6 t.nr (Normalization Temperature in ºC) When Au are adj
12 performa ce in extreme sample ranges. Use this setup function to eter default is 1.0 to match the nchange the cell constant if necessary. Minclud
13 7. CALCULATING TDS CONVERSION FACTOR You can calibrate TDS using the value of the calibration standard solution at a standard temperature such a
8. CALCULATING TEMPERATURE COEFFICIENTS To determine the temperature coefficient of your sample solution use this formula: Where: tc = Temperatur
15 9. REPLACEMENTS AND ACCESSORIES Part number Ordering Code Item Description Eutech Instruments Oakton Instruments CON 700 Benchtop with electrod
10. TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION No display Main power not switched on. AC Adapter socket not inserted properly. a) Switch on the p
17 11. SPECIFICATIONS Conductivity Range 0 to 20.00, 200.0, 2000 µS 0 to 20.00, 200.0 mS TDS Range (0.5 TDS factor) 0 to 10.00, 100.0, 1000 ppm
18 12. WARRANTY This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from
For more information on our products, please contact our channel partner or visit our websites listed below: Oakton Instruments 625 E Bunker
TABLE OF CONTENTS 1. INTRODUCTION... 1 2. GETTING STARTED...
6.12 CELL (Nominal Cell Constant) ...11 6.13 P4.0 rSt (Reset) ...
1 1. INTRODUCTION Thank you for purchasing our Con 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The
2. GETTING STARTED 2.1 Keypad Functions Powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used.
2.2 LCD Annunciators 2.3 Meter Connections CON/Temp 8-pin DIN connection for 2-cell Con/TDS/Temp electrode DC Power supply 3
3. ELECTRODE INFORMATION The CON700 includes an electrode with a nominal cell constant of k = 1.0, built-in temperature sensor, and 1 meter cable. Th
5 The factory default is automatic conductivity calibration. See Section 6.10 to change this setting. There is no automatic calibration value availa